10.18805/IJAR.BF-2095Cite article - Ünal Yavuz, Kerem Yener, Mehmet Salih Karadağ, Kübra DİKMEN İLGİNOĞLU, Mehmet Sıdık HURMA, Ali HAYAT, Mustafa ÖZEN. (2026). In vitro Biomechanical Testing of Polyurethane External Skeletal Fixator as a Novel Polymer in Fracture Treatment. Indian Journal of Animal Research. 60(10): 1800–1807. doi: 10.18805/IJAR.BF-2095
Background: To determine the temperature, time and biomechanical properties of free-form polyurethane external skeletal fixation (PUESF) and compare them with acrylic external skeletal fixation (ACESF).
Methods: Three different configurations (uniplanar, multiplanar and circular) (n=4 for each configuration) were subjected to compression and bending (n=24 PUESF, n=24 ACESF) and only the uniplanar configuration was subjected to pullout testing (n=4 PUESF, n=4 ACESF). The temperatures of the connecting bar and K-wire were recorded when creating the multiplanar and circular configurations. All the configurations were subjected to axial and bending tests and the uniplanar configuration was subjected to pullout tests.
Result: Regarding the connecting bar temperature and the temperature delivered to the K-wires, the PUESF group reached maximum temperatures of 75.7°C and 38.1°C at 3 minutes and the ACESF group reached maximum temperatures of 87.6°C and 60°C at 9 minutes, respectively (p<0.001). Under compression force, the PUESF group showed similar strength to the ACESF group in the uniplanar and circular configurations (p>0.05), while it showed less strength than the ACESF group in the multiplanar configuration (p<0.05). The PUESF group showed similar strength to the ACESF group in all three configurations under the bending force (p>0.05). The ACESF group showed approximately 2 times as much pullout strength as that of the PUESF group (p<0.05). Polyurethane showed acceptable mechanical performance compared to acrylic in several test configurations. Polyurethane can be used in clinical trials for free-form external skeletal fixations (ESFs).
Distinguished scholars and subject experts committed to editorial excellence, scientific integrity and rigorous peer review.
View Full Board